Connections from the SEII to a 3rd party pulse processor are typically made through the external Scanning Interface Unit's DB-25 rear-panel connector labelled PHA Logic I/O. In certain circumstances, a direct connection to the DB-37 connector on the back of a SEII board is allowed. Shown below are the connections for the 4560 and 4561 pulse processors.
SEII DB37 SIU DB25 4560/4561
(ADC Busy) 4/22 --->> 2/15 --->> 5/6 - Start of Conversion
(Pulse Reject) 5/23 <<--- 3/16 <<--- 4 - Reject
(Live Time) 3/21 <<--- 4/17 <<--- 2 - DT
(Fast Channel) 2/20 <<--- 6/19 <<--- 1 - C Rate
(PHA Input) 6/24 <<--- BNC <<--- BNC - Output
|
Special Notes: